NC TREK CORONARY DILATATION CATHETER

MDR report key: 3053016 · Received April 11, 2013

Report

Report Number
2024168-2013-02197
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 20, 2013
Report Date
March 20, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110134
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED SEPARATION WAS CONFIRMED. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. IT SHOULD BE NOTED THAT THE INSTRUCTION FOR USE STATES: IF RESISTANCE IS FELT, DETERMINE THE CAUSE BEFORE PROCEEDING. CONTINUING TO ADVANCE OR RETRACT THE CATHETER WHILE UNDER RESISTANCE MAY RESULT IN DAMAGE TO THE VESSELS AND / OR DAMAGE / SEPARATION OF THE CATHETER.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RETURNED FOR EVALUATION. THE INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS PROXIMAL RIGHT CORONARY ARTERY. WHILE INSERTING A 3.0 X 15 MM NC TREK, THERE WAS FORCE APPLIED AND THE PROXIMAL SHAFT (SNAPPED) SEPARATED. IT WAS EASILY REMOVED FROM THE ANATOMY AND ANOTHER BALLOON CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155099 NC TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 21214G1

Patients

Seq Age Sex Outcome Treatment
1 73 YR