FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY SPATULA INSTRUMENT

MDR report key: 3052943 · Received April 11, 2013

Report

Report Number
2955842-2013-01197
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
December 13, 2012
Report Date
March 13, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED A PIECE OF THE CERAMIC SLEEVE WAS BROKEN OFF. THE BROKEN PIECE WAS NOT RETURNED. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENT TIP LOOKED LIKE IT HAD SOME BUILD UP RESIDUE LEAVING IT DISCOLORED. AN ELECTRICAL CONTINUITY TEST PASSED. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI S SURGICAL PROCEDURE THE CAUTERY SEAL WAS BROKEN ON THE PERMANENT CAUTERY SPATULA INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156221 PERMANENT CAUTERY SPATULA INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420184-05 M10100928 211

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS & ACCESSORIES ESU