FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3052885 · Received April 2, 2013

Report

Report Number
1627487-2013-13494
Event Type
Injury
Date Received
April 2, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA). EVAL RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-13495. IT WAS REPORTED THE PATIENT WAS EXPERIENCING POCKET HEATING WHILE CHARGING HER SCS SYSTEM. THE PATIENT STATED HER SYSTEM IS WORKING VERY WELL FOR HER. A NEW CHARGER WAS SENT TO THE PATIENT. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIV, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135459 CHARGING SYSTEM SCS IPG CHARGER LGW ST. JUDE MEDICAL - NEUROMODULATION 3701 171649

Patients

Seq Age Sex Outcome Treatment
1 43 YR Required Intervention IMPLANT DATE:| SCS LEAD: MODEL 3186| SCS ANCHOR: MODEL 1192| SCS LEAD: MODEL 3186| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| IMPLANT DATE:| IMPLANT DATE: