FDA Adverse Event
Malfunction
Summary report: N
TENACULUM FORCEPS INSTRUMENT
MDR report key: 3052870
·
Received April 11, 2013
Report
- Report Number
- 2955842-2013-01195
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 12, 2013
- Report Date
- March 12, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THAT ONE OF THE PITCH CABLES WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT CONTAINING THE CRIMP STILL REMAINED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
PRIOR TO STARTING A DA VINCI SI PROCEDURE, THE USER FACILITY REPORTEDLY IDENTIFIED A BROKEN WIRE ON THE TENACULUMS FORCEPS INSTRUMENT'S WRIST. THE INSTRUMENT WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155939 | TENACULUM FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420207-06 | M10120801 873 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |