FDA Adverse Event Malfunction Summary report: N

TENACULUM FORCEPS INSTRUMENT

MDR report key: 3052870 · Received April 11, 2013

Report

Report Number
2955842-2013-01195
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 12, 2013
Report Date
March 12, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THAT ONE OF THE PITCH CABLES WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT CONTAINING THE CRIMP STILL REMAINED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

PRIOR TO STARTING A DA VINCI SI PROCEDURE, THE USER FACILITY REPORTEDLY IDENTIFIED A BROKEN WIRE ON THE TENACULUMS FORCEPS INSTRUMENT'S WRIST. THE INSTRUMENT WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155939 TENACULUM FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420207-06 M10120801 873

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES