FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 3052845 · Received April 11, 2013

Report

Report Number
2955842-2013-01194
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 12, 2013
Report Date
March 12, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE DAMAGED INSULATION ON THE INSTRUMENT. ENGINEERING FOUND THAT THE DISTAL END OF THE INSTRUMENT'S MAIN TUBE EXHIBITED VARIOUS SCRATCH MARKS AND GOUGE MARKS WITH LIGHT MATERIAL REMOVAL. ENGINEERING CONCLUDED THAT THE DAMAGE WAS MOST LIKELY CAUSED BY MISHANDLING (EXCESS FORCE CONTACT ON THE MAIN TUBE SURFACE). THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES:GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

PRIOR TO STARTING A DA VINCI SI PROCEDURE, THE USER FACILITY REPORTEDLY IDENTIFIED DAMAGED INSULATION ON A MONOPOLAR CAUTERY INSTRUMENT. THE INSTRUMENT WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156720 MONOPOLAR CAUTERY INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420142-04 S10110615 565

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES, ESU