FDA Adverse Event Injury Summary report: N

OMNILINK ELITE PERIPHERAL STENT SYSTEM

MDR report key: 3052843 · Received April 11, 2013

Report

Report Number
2024168-2013-02191
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 18, 2013
Report Date
March 18, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIO
PMA / PMN Number
P110043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PROCEDURE, AN 8.0MM X 19MM X 135CM OTW OMNILINK ELITE VASCULAR BALLOON-EXPANDABLE STENT SYSTEM WAS INTRODUCED VIA FEMORAL ACCESS, THEN ADVANCED TOWARD A CALCIFIED LESION IN THE NON-TORTUOUS RIGHT COMMON ILIAC ARTERY, HOWEVER, THE OMNILINK ELITE FAILED TO CROSS THE LESION. WHILE WITHDRAWING THE OMNILINK ELITE SYSTEM, THE STENT DISLODGED FROM THE BALLOON AND BECAME LODGED IN THE SUBCUTANEOUS TISSUE. THE STENT WAS REMOVED FROM THE SKIN SURFACE ACCESS SITE. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO OCCURRENCE OF A CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154320 OMNILINK ELITE PERIPHERAL STENT SYSTEM PERIPHERAL STENT SYSTEM NIO AV-TEMECULA-CT 2102241

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention