FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3052785 · Received April 11, 2013

Report

Report Number
2124215-2013-05591
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 6, 2013
Report Date
April 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE AND LEAD WERE SUCCESSFULLY EXPLANTED AND REPLACED. THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS CONFIRMED THE INNER CONDUCTOR COIL WAS FRACTURED 20 CENTIMETERS (CM) FROM THE TERMINAL PIN. DETAILED ANALYSIS OF THE FRACTURE SITE DETERMINED THE CONDUCTOR COIL WAS FRACTURED DUE TO CYCLIC FATIGUE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED NOISE. THE NOISE WAS OVERSENSED RESULTING IN INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND MULTIPLE SHOCKS. THE NOISE WAS ABLE TO BE REPRODUCED WITH MANIPULATION OF THE DEVICE HEADER. THE LOCAL FIELD REPRESENTATIVE CONTACTED BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) AND INQUIRED ABOUT ADVISORY STATUS OF THE DEVICE. TS DISCUSSED THAT THIS DEVICE IS INCLUDED IN THE (B)(6) 2009 SUBPECTORAL IMPLANT 2009 ADVISORY POPULATION. THE LOCAL FIELD REPRESENTATIVE INDICATED THAT THE DEVICE WAS NOT IMPLANTED SUBPECTORAL. AN INVASIVE PROCEDURE WAS PERFORMED. THE POCKET WAS OPENED AND A FRACTURE WAS OBSERVED IN THE LEAD. NOISE WAS PRODUCED WHEN THE LEAD WAS MANIPULATED NEAR THE FRACTURE AND ALSO NEAR THE DEVICE HEADER. A HEADER ISSUE WAS UNABLE TO BE RULED OUT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155651 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0185

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R 4549| 0185| N119| 4474| 4195