ALTRUA
Report
- Report Number
- 2124215-2013-04348
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- February 27, 2013
- Report Date
- April 17, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- PMA / PMN Number
- P970003
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE DEVICE REACHED END OF LIFE (EOL) AFTER ONLY THREE YEARS. SUBSEQUENTLY THE DEVICE WAS EXPLANTED DUE TO CONCERNS OVER PREMATURE BATTERY DEPLETION. DURING THE REVISION PROCEDURE BOTH THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE EXPLANTED DUE TO DISLODGEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 156376 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Hospitalization| L| R | K173| 4088| 4087| S603| 4137| 4136 |