FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 3052766 · Received April 11, 2013

Report

Report Number
2124215-2013-04348
Event Type
Injury
Date Received
April 11, 2013
Date of Event
February 27, 2013
Report Date
April 17, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
PMA / PMN Number
P970003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE DEVICE REACHED END OF LIFE (EOL) AFTER ONLY THREE YEARS. SUBSEQUENTLY THE DEVICE WAS EXPLANTED DUE TO CONCERNS OVER PREMATURE BATTERY DEPLETION. DURING THE REVISION PROCEDURE BOTH THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE EXPLANTED DUE TO DISLODGEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156376 ALTRUA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND S603

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| L| R K173| 4088| 4087| S603| 4137| 4136