FDA Adverse Event Malfunction Summary report: N

ALTRUA

MDR report key: 3052744 · Received April 11, 2013

Report

Report Number
2124215-2013-04770
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 1, 2013
Report Date
April 3, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A COPY OF DEVICE MEMORY WAS RECEIVED FOR ANALYSIS. REVIEW OF DEVICE MEMORY SHOWED NO HARDWARE RESETS, HOWEVER, THE MEMORY STATUS INDICATED THAT AN EXCECUTABLE RAM BANK FAILED, WHICH MAY CAUSE THE DEVICE TO REVERT TO RESET-VVI. ONCE THIS IS DECLARED AND SEVERAL 11 HOUR BATTERY MEASUREMENTS HAVE COMPLETED, THE GAS GUAGE AND LONGEVITY REMAINING SHOULD BE UPDATED WITH THE APPROPRIATE DATA. ADDITIONALLY, THE LOCAL FIELD REPRESENTATIVE INDICATED THAT ALL DEVICE SETTINGS WERE RETURNED TO NORMAL. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE IN-CLINIC FOLLOW UP, A POWER ON RESET MESSAGE WAS OBSERVED UPON INTERROGATION OF THIS PACEMAKER. THE DEVICE WAS PACING VVI 65 AND THE GAS GAUGE SHOWED FULL WITH A MAGNET RATE OF 100, HOWEVER, SHOWED A LONGEVITY OF LESS THAN 0.5 YEARS REMAINING. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THAT ALL PARAMETERS WOULD NEED TO BE PROGRAMMED BACK TO THE DESIRED SETTINGS AND THAT THE GAS GAUGE WOULD LIKELY NEED 5 MORE MEASUREMENTS TO BE ACCURATE. ADDITIONALLY, TECHNICAL SERVICES DISCUSSED PERFORMING A SAVE ALL TO DISK FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155732 ALTRUA NVZ GUIDANT CRM CLONMEL IRELAND S606

Patients

Seq Age Sex Outcome Treatment
1 81 YR S606| 4076