VITALITY 2
Report
- Report Number
- 2124215-2013-05522
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- September 30, 2012
- Report Date
- May 2, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. ANALYSIS WAS UNABLE TO CONFIRM THE ALLEGATION FROM THE FIELD.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE DEVICE INTERROGATION, IT WAS NOTED THAT THIS DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO AN EXTENDED CHARGE TIME THAT WAS OUTSIDE OF THE EXTENDED CHARGE TIME LIMIT. SUBSEQUENTLY THE DEVICE WAS EXPLANTED OVER TWO WEEKS LATER. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154771 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Hospitalization| L| R | 1720| T175| 0075| 1793 |