ALTRUA
Report
- Report Number
- 2124215-2013-04123
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- February 27, 2013
- Report Date
- August 20, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. THE DEVICE WAS THEN SUBJECTED TO A SERIES OF AUTOMATED DIAGNOSTIC TESTS THAT VERIFY THE PERFORMANCE OF THE PACING, SENSING AND RECORDING FUNCTIONS OF THE DEVICE. THE DEVICE PERFORMED NORMALLY THROUGHOUT ALL TESTS.
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
THE DEVICE WAS LATER EXPLANTED FOR AN UNSPECIFIED REASON AND RETURNED FOR RELIABILITY ANALYSIS.
ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS PROGRAMMED WITH NOMINAL SETTINGS, RESULTING IN BATTERY DEPLETION. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THIS CLINICAL OBSERVATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DECLARED ELECTIVE REPLACEMENT TIME (ERT) IN LESS THAN THREE YEARS FOR AN UNSPECIFIED REASON. ADDITIONAL INFORMATION WAS SOUGHT FROM THE LOCAL AREA SALES REPRESENTATIVE, HOWEVER, AT THIS TIME, ADDITIONAL INFORMATION WAS NOT AVAILABLE. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154678 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | S603| 4135| 4136 |