FDA Adverse Event Injury Summary report: N

FLEXTEND II

MDR report key: 3052643 · Received April 11, 2013

Report

Report Number
2124215-2013-03827
Event Type
Injury
Date Received
April 11, 2013
Date of Event
February 28, 2013
Report Date
May 21, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE AND LEADS REMAIN IMPLANTED. A FINAL REPORT WILL BE SENT AFTER INFORMATION IS RECEIVED OF THE REVISION PROCEDURE. IF THE PRODUCTS ARE RETURNED TO BOSTON SCIENTIFIC LABORATORY ANALYSIS WILL BE PERFORMED TO CONFIRM AND DETERMINE THE ROOT CAUSE OF THIS EVENT.

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE AND LEADS REMAIN IMPLANTED. NO ADDITIONAL INFORMATION WAS ABLE TO BE OBTAINED OF THE POTENTIAL REVISION PROCEDURE PLANNED DUE TO THE PATIENT'S HEART BLOCK. AT A PREVIOUS FOLLOW UP, THE PHYSICIAN STATED THAT ALL MEASUREMENTS WERE STABLE AND WITHIN NORMAL RANGE. THE PATIENT HAD A PROVOCATION TEST IN THE ATRIUM AND VENTRICLE OF A COMPLETE EXIT BLOCK. DUE TO NO RETURN ON THIS DEVICE AND LEAD A ROOT CAUSE CAN NOT BE DETERMINED AT THIS TIME FOR THE LOSS OF CAPTURE (LOC). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED AND AN AMENDED REPORT SUBMITTED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS DEVICE, EXPERIENCED HEART BLOCK AND WAS SUBSEQUENTLY HOSPITALIZED. THE DEVICE WAS INTERROGATED AND A LOSS OF CAPTURE (LOC) GREATER THAN TWO SECONDS WAS OBSERVED ON THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEAD. A REVISION PROCEDURE WAS SCHEDULED FOR THE NEAR FUTURE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156798 FLEXTEND II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4096

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R