ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-04400
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- February 28, 2013
- Report Date
- March 28, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE IMPLANT PROCEDURE, WHEN CLOSING THE POCKETS OVERSENSING OF A PREMATURE VENTRICULAR CONTRACTION (PVC) ON THE RIGHT VENTRICULAR (RV) CHANNEL THAT LEAD TO PACING INHIBITION WITH LESS THAN TWO SECONDS OF ASYSTOLE WAS OBSERVED. CLOSURE OF THE POCKET WAS STOPPED AND THE DEVICE WAS REMOVED FROM THE POCKET AND THE LEAD WAS REMOVED FROM THE HEADER. THE PHYSICIAN THEN EXPERIENCED DIFFICULTY RE-INSERTING THE LEAD INTO THE DEVICE. THUS A NEW RV LEAD WAS IMPLANTED. HOWEVER THE OVERSENSING OF NOISE WORSENED. SEVERAL ATTEMPTS WERE MADE AT RE-INSERTING THE LEAD WITHOUT ANY CHANGE. SUBSEQUENTLY A NEW DEVICE WAS IMPLANTED AND NO OTHER ISSUES WERE OBSERVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155374 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | MISMATCH| E163| 0185 |