FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 3052636 · Received April 11, 2013

Report

Report Number
2124215-2013-04400
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
February 28, 2013
Report Date
March 28, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE IMPLANT PROCEDURE, WHEN CLOSING THE POCKETS OVERSENSING OF A PREMATURE VENTRICULAR CONTRACTION (PVC) ON THE RIGHT VENTRICULAR (RV) CHANNEL THAT LEAD TO PACING INHIBITION WITH LESS THAN TWO SECONDS OF ASYSTOLE WAS OBSERVED. CLOSURE OF THE POCKET WAS STOPPED AND THE DEVICE WAS REMOVED FROM THE POCKET AND THE LEAD WAS REMOVED FROM THE HEADER. THE PHYSICIAN THEN EXPERIENCED DIFFICULTY RE-INSERTING THE LEAD INTO THE DEVICE. THUS A NEW RV LEAD WAS IMPLANTED. HOWEVER THE OVERSENSING OF NOISE WORSENED. SEVERAL ATTEMPTS WERE MADE AT RE-INSERTING THE LEAD WITHOUT ANY CHANGE. SUBSEQUENTLY A NEW DEVICE WAS IMPLANTED AND NO OTHER ISSUES WERE OBSERVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155374 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0185

Patients

Seq Age Sex Outcome Treatment
1 48 YR MISMATCH| E163| 0185