VITALITY 2
Report
- Report Number
- 2124215-2013-04027
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- August 12, 2012
- Report Date
- March 5, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS RETURNED WITH NO REPORTED PRODUCT PERFORMANCE ISSUES AND NO REPORTED ADVERSE PATIENT EFFECTS. INITIAL ANALYSIS COMPLETED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY DETERMINED THIS PRODUCT DID NOT MEET LONGEVITY EXPECTATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155303 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | T177| 0185 |