FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 3052607 · Received April 11, 2013

Report

Report Number
2124215-2013-03479
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
February 25, 2013
Report Date
March 5, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE DID NOT MEET A PORTION OF THE INITIAL DEVICE TESTING. ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS RETURNED FOR RELIABILITY ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154310 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 52 YR 4135| 0180| E143| T167