FDA Adverse Event Injury Summary report: N

INCEPTA

MDR report key: 3052599 · Received April 11, 2013

Report

Report Number
2124215-2013-04533
Event Type
Injury
Date Received
April 11, 2013
Date of Event
February 27, 2013
Report Date
June 4, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT DEFIBRILLATION THRESHOLD (DFT) TESTING WAS PERFORMED AND SYSTEM FUNCTION WAS OBSERVED TO BE NORMAL. THE PHYSICIAN WILL CONTINUE MONITORING. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION WAS RECEIVED THAT DEFIBRILLATION THRESHOLD (DFT) TESTING WAS PERFORMED AND SHOCK IMPEDANCE MEASUREMENTS WERE OBSERVED TO BE WITHIN NORMAL LIMITS. THE PHYSICIAN WILL CONTINUE MONITORING.

Additional Manufacturer Narrative · 1

(B)(4). THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE PHYSICIAN IS AWARE OF THE OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT AND IS CONTINUING TO MONITOR. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS CONTINUE TO BE OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED A HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155300 INCEPTA IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E162

Patients

Seq Age Sex Outcome Treatment
1 69 YR Life Threatening| R E162| 0292| 4135