FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3052578 · Received April 11, 2013

Report

Report Number
2124215-2013-04436
Event Type
Injury
Date Received
April 11, 2013
Date of Event
February 21, 2013
Report Date
December 16, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
LWS
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT AN INVASIVE PROCEDURE WAS PERFORMED. THE ICD WAS EXPLANTED AND REPLACED AND THIS LEAD REMAINS IMPLANTED AND IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT SHOCK IMPEDANCE MEASUREMENTS HAVE TRENDED HIGH FOR THIS DEVICE AND SINGLE COIL IMPLANTABLE DEFIBRILLATION LEAD. THE PHYSICIAN WILL CONTINUE TO MONITOR THE DEVICE/LEAD SYSTEM. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT SHOCK IMPEDANCE MEASUREMENTS HAVE TRENDED HIGH FOR THIS DEVICE AND SINGLE COIL IMPLANTABLE DEFIBRILLATION LEAD. THE PHYSICIAN WILL CONTINUE TO MONITOR THE DEVICE/LEAD SYSTEM. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

--

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS CONTINUE TO BE OBSERVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED A HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS. A REVIEW OF STORED DEVICE MEMORY REVEALED THAT SHOCK IMPEDANCE MEASUREMENTS HAD BEEN TRENDING IN THE UPPER 90 OHM TO LOWER 100 OHM RANGE WITH ONLY ONE GREATER THAN 125 OHMS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154257 ENDOTAK RELIANCE IMPLANTABLE LEAD LWS CPI - DEL CARIBE 0180

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| L| R 4469| 0180| E110