FLEXTEND
Report
- Report Number
- 2124215-2013-04405
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- February 25, 2013
- Report Date
- April 25, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. THE FIELD OBSERVATION COULD NOT BE CONFIRMED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE AND LOW PACING IMPEDANCES LESS THAN 200 OHMS. NO PACING INHIBITION WAS OBSERVED. SURGICAL INTERVENTION WAS PERFORMED AND THIS LEAD WAS EXPLANTED AND A NEW RV LEAD WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154923 | FLEXTEND | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Hospitalization| L| R | 4087| S603| 4088 |