FDA Adverse Event
Malfunction
Summary report: N
ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 3052454
·
Received April 11, 2013
Report
- Report Number
- 2250051-2013-00103
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- April 7, 2013
- Report Date
- April 11, 2013
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- MMH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE FE ARRIVED ON SITE AND INVESTIGATED THE INSTRUMENT FOR THE REPORTED ISSUE. THE FE INSPECTED THE X ARM AND IDENTIFIED A DAMAGED TIP CLAMP. FE REPLACED THE TIP CLAMP. THE FE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED THE SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND RETURNED TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).