FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3052383 · Received April 11, 2013

Report

Report Number
2032227-2013-01322
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 20, 2013
Report Date
March 20, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE INSULIN PUMP DID HAVE A SCRATCHED LCD WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH BLOOD GLUCOSE LEVELS GREATER THAN 600 MG/DL. IT WAS STATED THAT THE CUSTOMER EXPERIENCED FAINTING, ELEVATED HEART RATE, CHEST PRESSURE, DIFFICULTY BREATHING AND VOMITING FOR THREE DAYS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THERE WAS NO TUBING CLAMP FOR THE HIGH PRESSURE TEST. THE MOTHER STATED THAT SHE WOULD FEEL MORE COMFORTABLE HAVING THE INSULIN PUMP REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156256 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 30 YR Hospitalization