COBAS 6000 C501 MODULE
Report
- Report Number
- 1823260-2013-02205
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 26, 2013
- Report Date
- April 26, 2013
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JGS
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
THE CUSTOMER PROVIDED ADDITIONAL INFORMATION ON THE PATIENT IN QUESTION. THE INITIAL ION SELECTIVE ELECTRODE (ISE) CHLORIDE (CL) RESULT WAS 87.5 MMOL/L. THE SAMPLE WAS REPEATED ON COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4) AND GENERATED A REPEAT RESULT OF 105.4 MMOL/L. THE CUSTOMER DEEMED THE REPEAT RESULT TO BE THE CORRECT RESULT. THERE WAS NO ADVERSE EVENT.
THE INVESTIGATION WAS UNABLE TO DETERMINE A SPECIFIC ROOT CASE. IT WAS NOTED THAT A SAMPLE HANDLING ISSUE COULD CAUSE ERRONEOUS RESULTS, SINCE SUBSEQUENT MEASUREMENTS YEILDED PLAUSIBLE RESULTS.
THE CUSTOMER RECEIVED QUESTIONABLE RESULTS FOR ION SELECTIVE ELECTRODE (ISE) SODIUM (NA) AND ISE POTASSIUM (K) ON ONE PATIENT SAMPLE. IT WAS DETERMINED THE NA RESULT WAS ERRONEOUS. ALL RESULTS ARE IN MMOL/L. THE INITIAL NA RESULT WAS 117. THE SAMPLE WAS REPEATED ON COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4) AND GENERATED A RESULT OF 135. THE INITIAL RESULT HAD BEEN VERBALLY GIVEN TO THE NURSE. THE CUSTOMER DEEMED THE REPEAT RESULT TO BE THE CORRECT RESULT, AND THE NURSE WAS CALLED BACK WITH THE CORRECTED RESULT. THERE WAS NO ADVERSE EVENT. THE LOT NUMBER FOR THE NA ELECTRODE IN USE WAS NOT PROVIDED. THE CUSTOMER REFUSED A FIELD SERVICE DISPATCH FOR THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154693 | COBAS 6000 C501 MODULE | CLINICAL CHEMISTRY ANALYZER | JGS | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 035 YR |