FDA Adverse Event Malfunction Summary report: N

PROXIMATE PPH PROCEDURE SET

MDR report key: 3052314 · Received April 11, 2013

Report

Report Number
3005075853-2013-01722
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 9, 2013
Report Date
March 19, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051301
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STAPLED HEMORRHOIDECTOMY PROCEDURE WHEN THE SURGEON ATTEMPTED TO 'FIRE' THE INSTRUMENT THERE WAS NO RESPONSE FROM THE INSTRUMENT. IT DID NOT FIRE. ANOTHER INSTRUMENT WAS USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155965 PROXIMATE PPH PROCEDURE SET STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4CG58

Patients

Seq Age Sex Outcome Treatment
1