FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3052218 · Received April 4, 2013

Report

Report Number
1627487-2013-04412
Event Type
Injury
Date Received
April 4, 2013
Date of Event
January 17, 2013
Report Date
March 13, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS INCLUDED IN A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2013-04411. IT WAS REPORTED THAT PT WAS EXPERIENCING A PAINFUL BURNING SENSATION WHEN USING THE CHARGING SYSTEM TO CHARGE THE IPG. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139537 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3673633

Patients

Seq Age Sex Outcome Treatment
1 44 YR Required Intervention IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2)| SCS LEAD: MODEL 3186 (2)