FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT COMBO

MDR report key: 3052165 · Received April 11, 2013

Report

Report Number
2183996-2013-00650
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
April 8, 2013
Report Date
May 31, 2013
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE CHECK BUTTON DOES NOT RESPOND. THERE ARE PARTICLES INSIDE THE HOUSING OF THE CHECK BUTTON. THE PARTICLES ISOLATE THE AREA OF CONTACT INSIDE THE BUTTON HOUSING. DUE TO THIS PROBLEM THE CHECK BUTTON DOES NOT RESPOND AND IS WITHOUT FUNCTION.

Description of Event or Problem · 1

ON (B)(6) 2013, IT WAS REPORTED THAT OCCASIONALLY AFTER COMING INTO CONTACT WITH WATER THE CHECK BUTTON ON THE PATIENT'S INFUSION DEVICE WOULD NOT FUNCTION AND SOMETIMES THE DEVICE WOULD SHUT OFF BY ITSELF. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155205 ACCU-CHEK SPIRIT COMBO INSULIN INFUSION DEVICE LZG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 048 YR