FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 3052163 · Received April 11, 2013

Report

Report Number
2183996-2013-00648
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
April 3, 2013
Report Date
May 31, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE MOISTURE ENTERED THE INSULIN PUMP AND DESTROYS THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON DOES NOT RESPOND TO COMMANDS DUE TO THE CONTAMINATION INSIDE THE BUTTON.

Description of Event or Problem · 1

ON (B)(6) 2013, IT WAS REPORTED THAT THE SOFT COMPONENTS OF THE UP AND DOWN BUTTONS ON THE PATIENT'S INFUSION DEVICE WERE DETERIORATED AND TWO DAYS AGO THEY BEGAN TO ONLY FUNCTION INTERMITTENTLY. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156246 ACCU-CHEK SPIRIT INSULIN INFUSION DEVICE LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 028 YR