ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2013-00648
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- April 3, 2013
- Report Date
- May 31, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE MOISTURE ENTERED THE INSULIN PUMP AND DESTROYS THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON DOES NOT RESPOND TO COMMANDS DUE TO THE CONTAMINATION INSIDE THE BUTTON.
ON (B)(6) 2013, IT WAS REPORTED THAT THE SOFT COMPONENTS OF THE UP AND DOWN BUTTONS ON THE PATIENT'S INFUSION DEVICE WERE DETERIORATED AND TWO DAYS AGO THEY BEGAN TO ONLY FUNCTION INTERMITTENTLY. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 156246 | ACCU-CHEK SPIRIT | INSULIN INFUSION DEVICE | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 028 YR |