FDA Adverse Event Injury Summary report: N

MICRO CHPV W RICKHAM UNITIZED

MDR report key: 3052161 · Received April 11, 2013

Report

Report Number
1226348-2013-15175
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 21, 2013
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
JXG
PMA / PMN Number
PK980778
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

IT WAS NOTED THAT THIS COMPLAINT COULD NOT BE CONFIRMED. THE DEVICE WAS VISUALLY AND FUNCTIONALLY TESTED AND WAS FOUND TO WORK IN ACCORDANCE WITH THE MANUFACTURING SPECIFICATIONS. THE PROBLEM REPORTED BY THE CUSTOMER COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THAT THIS DEVICE CONFORMED TO THE REQUIRED SPECIFICATIONS PRIOR TO DISTRIBUTIONS. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Additional Manufacturer Narrative · 1

CODMAN HAS REQUESTED RETURN OF THE VALVE FOR EVALUATION. HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE ACCUMULATIONS OF BIOLOGICAL DEBRIS WITHIN THE VALVES OR CATHETERS, WHICH HAVE RESULTED IN BLOCKAGES WITHIN THE DEVICES, RESTRICTING FLOW. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CLOSED AT THIS TIME.

Description of Event or Problem · 1

THE AFFILIATE REPORTED THAT AFTER IMPLANTATION, IT WAS NOTED THAT FLUID DID NOT FLOW THROUGH THE DEVICE. AS A RESULT, THE DEVICE WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154851 MICRO CHPV W RICKHAM UNITIZED CNS SHUNT JXG CODMAN & SHURTLEFF, INC. CLFDMT

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention