FDA Adverse Event Injury Summary report: N

CERTAS INLIN VLV SPHN/UNIT CAT

MDR report key: 3052144 · Received April 11, 2013

Report

Report Number
1226348-2013-15178
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 15, 2013
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
JXG
PMA / PMN Number
PK112156
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

IT WAS NOTED THAT THIS COMPLAINT COULD NOT BE CONFIRMED. THE DEVICE WAS VISUALLY AND FUNCTIONALLY TESTED AND WAS FOUND TO WORK IN ACCORDANCE WITH THE MANUFACTURING SPECIFICATIONS. THE PROBLEM REPORTED BY THE CUSTOMER COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. A REVIEW OF THE DEVICE HISTORY RECORDS COULD NOT BE CONDUCTED AS THE LOT/SERIAL NUMBER HAS NOT BEEN PROVIDED. BASED ON THE RESULTS OF THIS INVESTIGATION, NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME, THIS COMPLAINT IS CLOSED.

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.

Description of Event or Problem · 1

THE AFFILIATE REPORTED THAT THE VALVE WAS NO LONGER ADJUSTABLE. AS A RESULT, THE DEVICE WAS REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155198 CERTAS INLIN VLV SPHN/UNIT CAT SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention