PROMUS ELEMENT ¿ LONG
Report
- Report Number
- 2134265-2013-02208
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- December 29, 2012
- Report Date
- March 15, 2013
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS RECEIVED FOR ANALYSIS AND VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE FOUND THAT THE STENT HAD MOVED DISTALLY ON THE BALLOON. THE STENT WAS ALSO DAMAGED AND BUNCHED AT THE PROXIMAL EDGE AND THERE WERE KINKS ALONG THE ENTIRE LENGTH OF THE HYPOTUBE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED. THE BALLOON WAS INFLATED WITH NO ISSUES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2013. SAME CASE AS MDR#: 2134265-2013-02209. IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, THE STENT DELIVERY BALLOON FAILED TO INFLATE. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 80% DE NOVO, CONCENTRIC, 2.75X33MM TARGET LESION WAS LOCATED IN THE NON-CALCIFIED AND MILDLY TORTUOUS LEFT ANTERIOR DESCENDING AND LEFT CIRCUMFLEX ARTERIES. THE LESION WAS PREDILATED WITH A 2.5X15 NON-BSC BALLOON CATHETER WHICH REDUCED THE STENOSIS TO 50%. A 2.75 X 38MM PROMUS ELEMENT MR STENT WAS ADVANCED TO THE LEFT ANTERIOR DESCENDING LESION; HOWEVER, THE PROMUS ELEMENT WAS UNABLE TO INFLATE. ANOTHER 2.75 X 38MM PROMUS ELEMENT MR STENT WAS ADVANCED TO THE TARGET LESION WHICH WAS ALSO UNABLE TO INFLATE. CONTRAST MEDIA AND RATIO IN THE INFLATION DEVICE WAS 75%. BOTH PROMUS ELEMENT STENT DEVICES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT BRAND STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT MOVED ON BALLOON AND STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154817 | PROMUS ELEMENT ¿ LONG | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911338270 | 0014561553 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | GUIDE WIRE: WHISPER| BALLOON CATH: SPRINTER 2.5 X 15| GUIDE CATHETER: 6FR XB 3.5 |