FDA Adverse Event Malfunction Summary report: N

INSERTER FOR TI ELASTIC NAILS

MDR report key: 3052137 · Received April 11, 2013

Report

Report Number
8030965-2013-10721
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
June 1, 2012
Report Date
June 2, 2012
Manufacturer
SYNTHES GMBH
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED.

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4). ORIGINAL AWARENESS DATE IS (B)(4) 2012.

Description of Event or Problem · 1

THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ROUTINE PLANNED ELASTIC NAIL REMOVAL, THE SURGEON WAS REPORTEDLY TRYING TO REMOVE THE TI ELASTIC NAIL. AS THE SURGEON WAS HAMMERING, THE THREADED END PORTION OF THE 359.218 EXTENSION BROKE OFF AND THE THREADS WERE STILL ATTACHED IN THE CHUCK 359.219. THE SURGEON WAS ABLE TO COMPLETE THE PROCEDURE WITHOUT USING OTHER INSTRUMENTATION. THE CONSULTANT PROVIDED LOANER INSTRUMENTS FROM THE METRO OFFICE TO USE IF NEEDED BEFORE REPLACEMENTS ARRIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156146 INSERTER FOR TI ELASTIC NAILS LXH SYNTHES GMBH 3098634

Patients

Seq Age Sex Outcome Treatment
1