FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3052048 · Received April 11, 2013

Report

Report Number
2531779-2013-04036
Event Type
Malfunction
Date Received
April 11, 2013
Report Date
March 15, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: NO INSULIN DELIVERY DEFECT WAS FOUND WITH THE PUMP. DISPLAY WAS FADED AND PINK. REMOVED DISPLAYED AND PLACED NEW GOOD DISPLAY IN. NEW DISPLAY CONTRAST RETURNED TO NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED THE DISPLAY SCREEN WAS DIFFICULT TO READ. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156713 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 41 YR