FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 3051980 · Received April 11, 2013

Report

Report Number
3008382007-2013-07574
Event Type
Malfunction
Date Received
April 11, 2013
Report Date
March 18, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP #1 ((B)(4) 2013) - DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED TO LFS AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(6) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, ANOTHER FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

THE METER/TEST STRIPS ASSOCIATED WITH THIS COMPLAINT HAS BEEN RETURNED TO LIFESCAN; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR 2 STRIP ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155053 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3359855

Patients

Seq Age Sex Outcome Treatment
1 7 YR