FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 3051957 · Received April 11, 2013

Report

Report Number
3008382007-2013-07525
Event Type
Malfunction
Date Received
April 11, 2013
Report Date
March 16, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(4) 2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER WAS FOUND TO HAVE A HIGH STANDBY CURRENT DUE TO DIODE D6 AND D7 WERE SOLDER. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING BATTERY INDICATOR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155142 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3275876

Patients

Seq Age Sex Outcome Treatment
1