FDA Adverse Event Malfunction Summary report: N

OPTIFLOW JUNIOR NASAL CANNULA

MDR report key: 3051873 · Received April 10, 2013

Report

Report Number
9611451-2013-00258
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
November 6, 2012
Report Date
November 9, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
CAT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT OPT316 OPTIFLOW JUNIOR NASAL CANNULA WAS RETURNED TO FPH IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE LEFT TUBE WAS DETACHED FROM THE CANNULA, CONFIRMING THE REPORTED FAULT. THE SPRING WAS EVENLY DISTRIBUTED ALONG ITS LENGTH AND NO DAMAGE WAS OBSERVED. AN ADHESIVE WAS PRESENT AT THE EXTERNAL SURFACE OF THE TUBE THAT WAS INSERTED INTO THE CANNULA. A LOT CHECK WAS NOT PERFORMED AS LOT INFORMATION WAS NOT PROVIDED. CONCLUSION: THE DAMAGE WAS CAUSED BY THE BABY PULLING THE TUBE AWAY FROM THE CANNULA, AS REPORTED BY THE NURSE AT THE HOSPITAL. ALL OPTIFLOW JUNIOR NASAL CANNULAS ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE LEAVING THE PRODUCTION LINE, AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE OPT316 OPTIFLOW JUNIOR NASAL CANNULA STATE THE FOLLOWING: "CHECK ALL CONNECTIONS, CAPS AND/OR PLUGS ARE TIGHT BEFORE USE." "PATIENT MONITORING IS RECOMMENDED."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE TUBE OF AN OPT316 OPTIFLOW JUNIOR NASAL CANNULA BECAME DISCONNECTED FROM ITS CANNULA AS THE BABY "RIPPED IT APART". IT WAS FURTHER REPORTED THAT THE COILED WIRE CAME OUT FROM THE TUBE. THERE WERE NO CLINICAL IMPLICATIONS REPORTED AS THE "BABY WAS CRYING AND MOVING A LOT, AND THUS ALREADY HAD HIGH SATURATION".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152234 OPTIFLOW JUNIOR NASAL CANNULA CAT CAT FISHER & PAYKEL HEALTHCARE LIMITED OPT316 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1