ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00259
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 12, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE WHERE IT WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED A HOLE APPROXIMATELY 20CM FROM THE PROXIMAL CONNECTOR ON THE EVAQUA EXPIRATORY LIMB OF THE RETURNED RT340 BREATHING CIRCUIT. A LOT CHECK COULD NOT BE PERFORMED AS A LOT NUMBER WAS NOT PROVIDED. CONCLUSION: BASED ON THE INSPECTION OF THE DAMAGE, IT IS LIKELY THAT THE EVAQUA EXPIRATORY LIMB WAS PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO BEING RELEASED FOR DISTRIBUTION. IN ADDITION, TUBING WEIGHT AND BOND STRENGTH TSTS ARE PERFORMED EVERY 15 MINUTES. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. THE KEY DIFFERENCE BETWEEN THE EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOUR FROM EXPIRED VENTILATORY GAS TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND NON-FPH CIRCUIT HANGERS. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT AN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STARTED TO LEAK AFTER TWO DAYS OF USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT AN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STARTED TO LEAK AFTER TWO DAYS OF USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 152199 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DRAGER EVITA 2 VENTILATOR| MR850 RESPIRATORY HUMIDIFIER| DRAGER EVITA 2 VENTILATOR| MR850 RESPIRATORY HUMIDIFIER |