FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3051705 · Received April 10, 2013

Report

Report Number
3004209178-2013-92500
Event Type
Injury
Date Received
April 10, 2013
Date of Event
March 28, 2013
Report Date
March 28, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO LOW BLOOD GLUCOSE OF 20MG/DL. THE CUSTOMER STATED THAT THE INSULIN PUMP HAS A CRACK UNDER THE RESERVOIR WINDOW, AND IT GOES FROM THE LEFT CORNER OF THE WINDOW TO BETWEEN THE ESCAPE AND ACTIVATE BUTTON. THE CALLER STATED THAT SHE IS NOT SURE IF THE DEVICE WAS BUMPED ON THE DESK AT WORK. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153931 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 48 YR Hospitalization