FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3051684
·
Received April 10, 2013
Report
- Report Number
- 3004209178-2013-92491
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 27, 2013
- Report Date
- April 5, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 233MG/DL. THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED, AND THE DRIVE SUPPORT CAP WAS FLUSH. THE CALLER ALSO STATED THAT THE DEVICE WAS STUCK ON THE PRIME LOOP. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 152284 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |