COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-00611
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 13, 2013
- Report Date
- March 13, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
RESULTS: SCATTER MASK REPLACEMENT, VCS (VOLUME, CONDUCTIVITY, SCATTER) OPTIMIZATION AND CHEMICAL BALANCING.
THE CUSTOMER REPORTED OBTAINING HIGH EO (EOSINOPHILS) RESULTS FOR 2 PATIENT SAMPLES INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER INDICATED THAT MANUAL SMEARS OF THE SAMPLES REVEALED NO EO PRESENT AND WAS CONSIDERED CORRECT. ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. DATA REVIEW OF THE 2 SAMPLE RESULTS PROVIDED INDICATES THAT THE LH 750 GAVE LOWER NE% (NEUTROPHIL PERCENTAGE) AND HIGHER EO RESULTS WITHOUT INSTRUMENT-GENERATED FLAGS WHEN COMPARED TO RESULTS FROM MANUAL SMEARS. THE SAME SAMPLES WERE RERUN THE FOLLOWING DAY AND RESULTS CORRELATED TO THE MANUAL SMEARS AND WERE CONSIDERED CORRECT. QUALITY CONTROLS (QC) RUN BEFORE THE INCIDENT RECOVERED WITHIN LIMITS AND THE INSTRUMENT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND OBSERVED THE ELEVATED EO RESULT. THE FSE REPLACED THE LASER, E-LYSE PUMP, SCATTER SENSOR, AND SCATTER MASK, SOL (SOLENOID) 60 AND 63, AND THE ACTUATORS FOR PINCH VALVE PV65 AND PV46B. THE FSE ALSO PERFORMED VCS (VOLUME, CONDUCTIVITY, SCATTER) OPTIMIZATION AND CHEMICAL BALANCING TO RESOLVE THE ERRONEOUSLY ELEVATED EO ISSUE. INSTRUMENT WAS VERIFIED PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 152186 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |