FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 3051583 · Received April 10, 2013

Report

Report Number
0001811755-2013-00748
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 19, 2013
Report Date
March 19, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE UPPER AND LOWER BEARING AND THE UPPER BEARING WAS SHATTERED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS BEING TESTED AT THE MANUFACTURER'S FACILITY WHEN TESTING WAS ABORTED DUE TO THE HANDPIECE WOBBLING AND SHAKING. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149318 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT DRILL, SURGICAL, ENT (ELECTRIC ERL STRYKER INSTRUMENTS-KALAMAZOO 11131

Patients

Seq Age Sex Outcome Treatment
1