FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3051581 · Received April 10, 2013

Report

Report Number
1416980-2013-09001
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
February 4, 2013
Report Date
March 21, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE. SECTION 12 "PROGRAMMING A PRESCRIPTION" IN 12.3.3 ON PAGE 12-6 GIVES THE WARNING THAT "IF YOU SET THE MINIMUM DRAIN VOLUME% TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD CAUSE AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 23:25:00. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 721ML, INDICATING THE HOME PATIENT (HP) DRAINED 721ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1100ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152011 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 65 YR