FDA Adverse Event Injury Summary report: N

CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS

MDR report key: 3051438 · Received April 10, 2013

Report

Report Number
2015691-2013-19784
Event Type
Injury
Date Received
April 10, 2013
Date of Event
February 25, 2013
Report Date
March 12, 2013
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S016
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNFORTUNATELY, THE EXPLANTED DEVICE WAS NOT RETURNED TO EDWARDS FOR ANALYSIS; THEREFORE, THE SOURCE OF THE REPORTED CALCIFICATION COULD NOT BE ASSESSED. ALTHOUGH THE ROOT CAUSE OF THIS EVENT CANNOT BE CONFIRMED, IT APPEARS THAT THE STENOSIS AND REGURGITATION WERE LIKELY CAUSED BY THE CALCIFICATION. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.

Description of Event or Problem · 1

EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE PROSTHETIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 5 YEARS, 7 MONTHS DUE TO MITRAL STENOSIS AND REGURGITATION WITH CALCIFICATION. UNFORTUNATELY, THE EXPLANTED DEVICE WAS DISCARDED BY THE HOSPITAL. THERE HAVE NOT BEEN ANY ALLEGATIONS OF A MALFUNCTION OR DEFICIENCY RELATED TO THE EXPLANTED VALVE. THE DEVICE WAS REPLACED WITH ANOTHER EDWARDS VALVE. NO OTHER DETAILS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148448 CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 6900P

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| R