FDA Adverse Event Malfunction Summary report: N

CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP

MDR report key: 3051407 · Received April 10, 2013

Report

Report Number
0001811755-2013-00742
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 14, 2013
Report Date
March 14, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CONSOLE PASSED ALL ELECTRICAL SAFETY TESTING AS WELL AS A FUNCTION INSPECTION. THE ONLY FAILURE FOUND WITH THE CONSOLE WAS THE PLASTIC HOUSING OF THE POWER ENTRY ASSEMBLY WAS CRACKED. THIS WOULD NOT CAUSE A HANDPIECE TO RUN WITHOUT ACTIVATION. A FOOTSWITCH WAS ALSO REPORTED AND EVALUATED FOR THIS COMPLAINT. THE FOOTSWITCH WAS FOUND TO HAVE A WEAK CONNECTION BETWEEN THE CABLE AND CONNECTOR; THIS FAILURE CAN CAUSE A HANDSWITCH TO RUN WITHOUT ACTIVATION.

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED. NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP WAS BEING USED WITH A TPS BI-DIRECTIONAL FOOTSWITCH WHEN IT CAUSED UNREPORTED AMOUNT OF CORE MICRO DRILLS TO RUN WITHOUT USER ACTIVATION DURING CASES. IN ONE SPECIFIC EVENT, IT CAUSED A BILATERAL ANKLE REPLACEMENT PROCEDURE TO BE CANCELLED HALFWAY THROUGH. ATTEMPTS TO OBTAIN FURTHER DETAIL REGARDING THESE EVENTS WERE NOT SUCCESSFUL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP WAS CAUSING AN UNREPORTED AMOUNT OF HANDPIECES TO RUN WITHOUT USER ACTIVATION DURING CASES. IN ONE SPECIFIC EVENT, IT CAUSED A BILATERAL ANKLE REPLACEMENT PROCEDURE TO BE CANCELED HALFWAY THROUGH. ATTEMPTS TO OBTAIN FURTHER DETAIL REGARDING THESE EVENTS WERE NOT SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149831 CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 TPS BI-DIRECTIONAL FOOTSWITCH (B)(4)| CORE MICRO DRILL (B)(4)