FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 3051292 · Received April 10, 2013

Report

Report Number
2182208-2013-01047
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
July 23, 2012
Report Date
July 23, 2012
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; PROGRAMMER STYLUS WAS NOT RESPONDING. DISPLAY BEZEL OVERLAY ASSEMBLY FOUND OUT OF ELECTRICAL SPECIFICATION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE STYLUS WAS NOT RESPONDING. THE PROGRAMMER WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153632 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2090

Patients

Seq Age Sex Outcome Treatment
1