FDA Adverse Event Injury Summary report: N

ELECTRODE, PACEMAKER, PERMANENT

MDR report key: 3051267 · Received April 10, 2013

Report

Report Number
2649622-2013-04810
Event Type
Injury
Date Received
April 10, 2013
Report Date
July 25, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND ON THE ATRIAL PACING LEAD WAS VARIABLE. HIGH AND LOW IMPEDANCES NOTED THROUGHOUT THE LEAD TREND DATA. LEAD WARNING RECORDED ON (B)(4) 2013.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076 IMPLANTABLE PACING LEAD 2005 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE AND NOISE. THEREFORE THE RA LEAD WAS CAPPED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152236 ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER