FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 3051263 · Received April 10, 2013

Report

Report Number
2182208-2013-01052
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
July 23, 2012
Report Date
July 23, 2012
Manufacturer
RICE CREEK MANUFACTURING
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; PROGRAMMER HAD BROKEN PINS INSIDE THE ELECTROGRAM OUTPUT CONNECTOR. ANALYSIS ALSO FOUND THE POWER CORD DOOR IS DAMAGED, HINGE BULLET COVER IS MISSING, AND HINGE PLATE SCREW ARE MISSING.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER HAD BROKEN PINS INSIDE THE ELECTROGRAM OUTPUT CONNECTOR. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152169 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MANUFACTURING 2090

Patients

Seq Age Sex Outcome Treatment
1