CONSULTA CRT-D
Report
- Report Number
- 3004209178-2013-05926
- Event Type
- Injury
- Date Received
- April 10, 2013
- Report Date
- July 23, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Removal / Correction Number
- Z-0111-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS EVENT INVOLVES A LEGAL CASE IN PROGRESS OR POTENTIAL LITIGATION. IF FOLLOW-UP WAS REQUIRED, EFFORTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. THE PROPRIETARY NATURE OF THE EVENT MAY AFFECT FOLLOW UP EFFORTS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT DEVICES: 3830 IMPLANTABLE PACING LEAD: (B)(6) 2009. 4195 LEFT VENTRICULAR (LV) LEAD: (B)(6) 2009. (B)(4).
THIS DEVICE WAS INCLUDED IN A FIELD ACTION, BUT PRODUCT ANALYSIS FOUND THAT THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. EVALUATION SUMMARY: ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
IT WAS REPORTED THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PREMATURE BATTERY DEPLETION. THE DEVICE WAS PROGRAMMED TO HIGHER OUTPUTS FOR HIGH LEFT VENTRICULAR THRESHOLDS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 151894 | CONSULTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D224TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00046 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD |