FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 3051200 · Received April 10, 2013

Report

Report Number
3004209178-2013-05926
Event Type
Injury
Date Received
April 10, 2013
Report Date
July 23, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT INVOLVES A LEGAL CASE IN PROGRESS OR POTENTIAL LITIGATION. IF FOLLOW-UP WAS REQUIRED, EFFORTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. THE PROPRIETARY NATURE OF THE EVENT MAY AFFECT FOLLOW UP EFFORTS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT DEVICES: 3830 IMPLANTABLE PACING LEAD: (B)(6) 2009. 4195 LEFT VENTRICULAR (LV) LEAD: (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN A FIELD ACTION, BUT PRODUCT ANALYSIS FOUND THAT THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. EVALUATION SUMMARY: ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PREMATURE BATTERY DEPLETION. THE DEVICE WAS PROGRAMMED TO HIGHER OUTPUTS FOR HIGH LEFT VENTRICULAR THRESHOLDS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151894 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00046 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD