SPRINT QUATTRO
Report
- Report Number
- 2649622-2013-04833
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- November 27, 2012
- Report Date
- December 5, 2012
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SY
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. THERE WAS AN ALERT FOR HIGH RESISTANCE/IMPEDANCE. OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCES ARE RECORDED ON (B)(4) 2012 (IMPLANT DATE) AND (B)(4) 2013. DEFIB ACTIVE CAN AND VENTRICULAR PACE IMPEDANCES ARE HIGH AT IMPLANT (254 AND 4047 OHM RESPECTIVELY); DEFIB ACTIVE CAN AND SUPERIOR VENA CAVA (SVC) IMPEDANCES RISE AGAIN THE WEEK ENDING (B)(4) 2013. CONCOMITANT PRODUCTS 5076 IMPLANTABLE PACING LEAD (B)(6) 2012. (B)(4).
IT WAS REPORTED THAT THERE WAS A LEAD INTEGRITY ALERT (LIA) FOR HIGH RIGHT VENTRICULAR (RV) AND SUPERIOR VENA CAVA (SVC) IMPEDANCE IMMEDIATELY AFTER IMPLANT. THE IMPEDANCE VALUES WERE WITHIN NORMAL RANGE THE NEXT DAY USING ANOTHER PROGRAMMER, HOWEVER, ANOTHER ALERT TRIGGERED TWO MONTHS LATER. THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 152959 | SPRINT QUATTRO | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | D284DRG IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |