FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3051167 · Received April 10, 2013

Report

Report Number
2649622-2013-04840
Event Type
Injury
Date Received
April 10, 2013
Report Date
December 4, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE PROXIMAL CONDUCTOR WAS FRACTURED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE WAS PRESENT ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD EXPERIENCED AN APPARENT FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152852 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R (B)(4) COMPEITITOR IMPLANTABLE CARDIOVERTER DEFIBR