FDA Adverse Event Malfunction Summary report: N

VIRTUOSO DR

MDR report key: 3051109 · Received April 10, 2013

Report

Report Number
3004209178-2013-05890
Event Type
Malfunction
Date Received
April 10, 2013
Report Date
January 14, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS REVEALED WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MINIMUM BATTERY VOLTAGE IS 3.0 TO 2.66 VOLTS BETWEEN (B)(6) 2012 AND (B)(6) 2013. RECOMMENDED REPLACEMENT TIME INDICATOR IS LESS THAN OR EQUAL TO 2.62 VOLTS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS REVEALED WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MINIMUM BATTERY VOLTAGE IS 3.0 TO 2.66 VOLTS BETWEEN (B)(4) 2012 AND (B)(4) 2013. RECOMMENDED REPLACEMENT TIME INDICATOR IS LESS THAN OR EQUAL TO 2.62 VOLTS. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD 2008 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE DEVICE WAS EXPLANTED AND WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED, THE DEVICE BATTERY LONGEVITY PERFORMANCE WAS NOT COMPENSATORY WITH THE PROGRAMMED PARAMETERS AND THERAPY USAGE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153828 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00070 YR 6947 IMPLANTABLE TACHY LEAD