FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 3050963 · Received April 10, 2013

Report

Report Number
2649622-2013-04796
Event Type
Injury
Date Received
April 10, 2013
Report Date
January 23, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 6947 IMPLANTABLE TACHY LEAD, (B)(6) 2012; IMPLANTABLE PACING LEAD, (B)(6) 2012. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PHRENIC NERVE STIMULATION WHEN BEING LV PACED AT HIGHER OUTPUTS. THE LEFT VENTRICULAR (LV) LEAD WAS REPROGRAMMED, AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151847 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Required Intervention D314TRM IMPLANTABLE DEFIBRILLATOR