FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3050950 · Received April 10, 2013

Report

Report Number
2182208-2013-01042
Event Type
Malfunction
Date Received
April 10, 2013
Report Date
February 27, 2013
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE DEVICE FAILED TELEMETRY TESTING DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED, THE PROGRAMMER RF (RADIO-FREQUENCY) HEAD WAS RETURNED BECAUSE IT WAS THE WRONG ITEM. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149767 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2067

Patients

Seq Age Sex Outcome Treatment
1